Syndesmosis fixation: analysis of shear stress via axial load on 3.5-mm and 4.5-mm quadricortical syndesmotic screws.

Hansen Matthew, Le Long, Wertheimer Stuart, Meyer Eric, Haut Roger

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

The effect of shear stress on a fixated distal syndesmosis of the ankle was evaluated with a servohydraulic materials-testing machine. Eighteen syndesmoses were fixated in a quadricortical technique using 3.5-mm cortical and 4.5-mm cortical stainless steel screws. A shear stress was applied via an axial load in an attempt to simulate weightbearing. The 4.5-mm quadricortical screws produced a higher yield load and peak load (484.3 +/- 93.8 N and 597.7 +/- 81.4 N) when compared with the 3.5-mm quadricortical syndesmotic screws (412.8 +/- 55 N, P = .033 and 511.2 +/- 64.4 N). These findings suggest that a larger diameter screw provides greater resistance to an applied shear stress at the distal syndesmosis.

Для цитирования: Hansen Matthew, Le Long, Wertheimer Stuart, Meyer Eric, Haut Roger. Syndesmosis fixation: analysis of shear stress via axial load on 3.5-mm and 4.5-mm quadricortical syndesmotic screws.. The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons. 2006. DOI: 10.1053/j.jfas.2005.12.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.