MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jim.2024.113627

Медицинская библиотека · Journal of immunological methods · 2024

Comparison of three methods for the detection of antibodies against muscle-specific kinase.

Luong Kyphuong, Lozier Bucky K, Novis Camille L, Smith Tammy L, Zuromski Lauren M, Peterson Lisa K

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To compare 3 different methods for the detection of antibodies against muscle-specific kinase (MuSK).

METHODS

MuSK antibody testing was performed in 237 serum samples by enzyme-linked immunosorbent assay (ELISA) and fixed cell-based assay (f-CBA-IFA). One hundred and forty-eight (148) of the sera had previously been tested by RIA during clinical testing: 47 MuSK antibody positive and 101 MuSK antibody negative. Of the MuSK RIA negative antibodies, 46 tested positive for other neural antibodies. Additionally, 89 sera were subsequently tested by all three methods: 70 healthy controls and 19 sera positive for other neural antibodies.

RESULTS

Qualitative inter-assay agreement based on tiered RIA values was 100% for results of 1.00 nmol/L or greater by both methods; 81% and 94% for results between 0.21 and 0.99 nmol/L by ELISA and f-CBA-IFA, respectively; and 0% for results of 0.04-0.20 nmol/L by both methods. Negative results showed 100% agreement between RIA and both ELISA and f-CBA-IFA (n = 55). None of the controls positive for other neural autoantibodies or healthy controls were positive in any assay.

CONCLUSION

Overall, excellent agreement was observed between the 3 methods used to detect antibodies against MuSK. Both the f-CBA-IFA and ELISA performed comparably to RIA and exhibited excellent overall accuracy for MuSK IgG detection, with the f-CBA-IFA demonstrating higher agreement between positive samples with the RIA than the ELISA without identifying false positives in the control samples. Advantages of non-radioactive methods for the detection of MuSK antibodies include reduced handling and disposal of hazardous materials, potential for automation and the reagents having a longer shelf-life, reducing costs associated with both workflow and lot validations. Thus, commercially available ELISA and transfected cell-based assays are viable alternatives to the traditional radioactive assay used for serologic determination of MuSK IgG.

Для цитирования: Luong Kyphuong, Lozier Bucky K, Novis Camille L, Smith Tammy L, Zuromski Lauren M, Peterson Lisa K. Comparison of three methods for the detection of antibodies against muscle-specific kinase.. Journal of immunological methods. 2024. DOI: 10.1016/j.jim.2024.113627.

Предметные рубрики MeSH

Подготовка и перевод публикации