DOI 10.1097/00000542-200606000-00010

Медицинская библиотека · Anesthesiology · 2006

Comparison of three disposable extraglottic airway devices in spontaneously breathing adults: the LMA-Unique, the Soft Seal laryngeal mask, and the Cobra perilaryngeal airway.

van Zundert André, Al-Shaikh Baha, Brimacombe Joseph, Koster Jurgen, Koning Dick, Mortier Eric P

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The authors compared three disposable extraglottic airway devices in spontaneously breathing anesthetized adults: the LMA-Unique (LMA-U; The Laryngeal Mask Company, San Diego, CA), the Soft Seal laryngeal mask (SS-LM; Portex Ltd., Hythe, United Kingdom), and the Cobra perilaryngeal airway (Cobra-PLA; Engineered Medical Systems, Inc. Indianapolis, IN).

METHODS

Three hundred twenty consecutive adults (American Society of Anesthesiologists physical status I-III; aged 18-80 yr) were randomly allocated for airway management with one of the three devices. Anesthesia was with fentanyl-propofol for induction and a sevoflurane-nitrous oxide-oxygen-fentanyl mixture for maintenance. Intraoperative data were collected by an unblinded observer about ease of insertion, effective airway time, oropharyngeal leak pressure, anatomical position (determined with a rigid endoscope), intracuff pressure changes, and airway trauma. Data were collected by a blinded observer about sore throat, dysphagia, and dysphonia 2 h after surgery.

RESULTS

Insertion was easier with the LMA-U and SS-LM than with the Cobra-PLA (P < 0.02), but the overall failure rates were similar. Effective airway times were similar among groups. Oropharyngeal leak pressure was lower with the LMA-U than with the SS-LM and Cobra-PLA (P < 0.001). Intracuff pressure increased during surgery with all extraglottic airway devices. Anatomical position was better with the Cobra-PLA than with the SS-LMA (P < 0.001) and better with the SS-LM than with LMA-U (P < 0.001). Blood staining was detected more frequently with the Cobra-PLA than with the LMA-U and SS-LM (P < 0.001), but there were no differences in airway morbidity.

CONCLUSION

The LMA-U and SS-LM are easier to insert and cause less trauma than the Cobra-PLA, but the Cobra-PLA has a more effective seal than the LMA-U and better endoscopically determined anatomical position than the LMA-U and SS-LM.

Для цитирования: van Zundert André, Al-Shaikh Baha, Brimacombe Joseph, Koster Jurgen, Koning Dick, Mortier Eric P. Comparison of three disposable extraglottic airway devices in spontaneously breathing adults: the LMA-Unique, the Soft Seal laryngeal mask, and the Cobra perilaryngeal airway.. Anesthesiology. 2006. DOI: 10.1097/00000542-200606000-00010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.