Analytical Comparison of Three Commercial Lysing Reagents for Full-Spectrum Flow Cytometric Immunophenotyping.

Huamanyalli-Villar Elizabeth, Moya-Naranjo Julia, Sanchez-Jacinto Billy, Espíritu-Antezano Julio, Rosales-Rimache Jaime

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Flow cytometry (FC) enables rapid identification of cell lineages based on their fluorescent and light-scattering properties, playing a critical role in diagnosing and monitoring oncohematological diseases. Standardization of procedures in spectral FC is essential for method validation studies.

METHODS

We designed a comparative study using 21 blood samples from healthy donors. Three commercial lysing solutions used in spectral FC immunophenotyping were evaluated. Samples were stained with a panel of eight antibody-conjugated fluorochromes and aliquoted for assessment with each lysing solution (Excellyse I, Excellyse Easy, and Tombo RBC Lysis Buffer), following the manufacturers' protocols. Data acquisition was performed using a full-spectrum flow cytometer.

RESULTS

An average cell loss of 31.4% was observed, with no statistically significant differences between lysing solutions, although Excellyse I showed lower cell loss under the no-wash protocol. This reagent also achieved superior resolution in distinguishing lymphocytes from debris. It demonstrated the lowest coefficients of variation in both scatter and fluorescence parameters of leukocyte populations, highlighting its ability to preserve the optical characteristics of stained cells. RBC-T performed well in specific channels such as APC and PE-Cy7, whereas Excellyse Easy produced a resolution comparable to Excellyse I, albeit with greater signal dispersion.

CONCLUSION

Excellyse I demonstrated the best overall performance, with greater cell preservation, clearer population resolution, and lower variability in scatter and fluorescence. Nonetheless, Excellyse Easy may still be considered a viable alternative.

Для цитирования: Huamanyalli-Villar Elizabeth, Moya-Naranjo Julia, Sanchez-Jacinto Billy, Espíritu-Antezano Julio, Rosales-Rimache Jaime. Analytical Comparison of Three Commercial Lysing Reagents for Full-Spectrum Flow Cytometric Immunophenotyping.. Journal of clinical laboratory analysis. 2025. DOI: 10.1002/jcla.70140.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.