Efecto de la ventilación mecánica en la variabilidad de la oximetría de pulso en el posoperatorio de cirugía cardiaca.
Аннотация
OBJECTIVE
To describe whether mechanical ventilation participates in the variability of pulse oximetry in subjects in critical conditions.
METHODS
Postoperative cardiac surgery subjects with and without mechanical ventilation were analized; their demographic variables, heart rate and arterial blood gases were defined. Variability was obtained from the comparison of arterial saturation (SaO2) and partial oxygen saturation (SpO2). They were obtained from blood gases, a pocket-sized Nonin pulse oximeter, and a Phillips monitor. It was analyzed with the Bland-Altman procedure.
RESULTS
260 subjects and 780 pairs of comparisons were studied. The mean difference (bias) with and without mechanical ventilation: SaO2 monitor 0.0 (6.3, -6.3) vs. 0.1 (6.3, -6.2); SaO2 Nonin 1.2 (7.7, -5.3) vs. 1.4 (7.2, -4.4); SpO2 Nonin - SpO2 monitor 1.2 (6.6, -4.2) vs. 1.3 (5.8, -3.2). The heart rate: auscultation monitor -0.5 (3.6, -4.5) vs. -0.1 (3, -3.3); auscultation Nonin -1.1 (7.4, -9.7) vs. -0.8 (4.2, -5.7); monitor - Nonin -0.6 (3, -4.3) vs. -0.6 (3, -4.3).
CONCLUSIONS
Mechanical ventilation does not increase heart rate or partial oxygen saturation variability in critically ill subject's post-cardiac surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.