Antimicrobial Envelopes for Cardiac Device Infection Prevention: A Meta-Analysis of Randomized and Matched Studies.

Prata Alonzo Armani, Cavalcante Luís Fernando Ferreira, Stutz Gabriela Blanco, Melamed Thalia, Lima Caio Rezende, Locke Andrew H, Brígido Alexandra Régia Dantas

СтатьяМетаанализОбзорПеревод в очереди

Аннотация

INTRODUCTION

Cardiac implantable electronic device (CIED)-associated infections are a significant procedural complication. Antibacterial envelopes have emerged as a potential strategy to mitigate this risk, though the evidence regarding their prognostic benefit remains inconclusive.

OBJECTIVES

We performed a meta-analysis of randomized controlled trials (RCTs) and propensity score-matched cohorts evaluating the efficacy of the adjunctive use of antimicrobial envelopes in preventing CIED-related infections compared to standard prophylactic measures alone.

METHODS

We searched through PubMed, Cochrane, and Embase databases. Our primary outcomes were (1) any CIED-related infection and (2) major CIED-related infection. We conducted a sensitivity analysis in high-risk patients.

RESULTS

A total of one RCT and eight observational studies were included, involving 78,587 patients, of whom 12,290 (15.6%) received antimicrobial envelopes. The overall risk ratio (RR) analysis showed no significant difference in the incidence of any CIED-related infections between the antimicrobial envelope and standard prophylaxis groups (RR 0.55; 95% CI 0.30-1.01; I2 = 75%). However, time-to-event analysis demonstrated that the use of antimicrobial envelopes was associated with a significantly lower risk of any CIED-related infection [hazard ratio (HR) 0.38; 95% CI 0.17-0.87; I2 = 67%] and major CIED-related infection (HR 0.54; 95% CI 0.39-0.76; I2 = 0%). Moreover, among high-risk patients, envelope implantation significantly reduced infection risk (RR 0.64; 95% CI 0.47-0.88; I2 = 22%).

CONCLUSION

Our analysis demonstrates that antibacterial envelopes significantly reduce the risk of CIED-related infections, particularly in high-risk patients. Although the additional cost remains a consideration, the time-to-event findings and benefit observed in selected high-risk populations may support their use in appropriately stratified patients.

Для цитирования: Prata Alonzo Armani, Cavalcante Luís Fernando Ferreira, Stutz Gabriela Blanco, Melamed Thalia, Lima Caio Rezende, Locke Andrew H, Brígido Alexandra Régia Dantas. Antimicrobial Envelopes for Cardiac Device Infection Prevention: A Meta-Analysis of Randomized and Matched Studies.. Pacing and clinical electrophysiology : PACE. 2025. DOI: 10.1111/pace.70108.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.