INTRODUCTION

We aim to evaluate the cyclic and torsional fatigue resistance of two rotary instruments, Hyflex CM 25/0.06 (HCM) (Coletene-Whaledent, Allstetten, Switzerland) and Aurum Blue (AB) 25/0.06 (Meta-Biomed, Republic of Korea).

METHODS AND MATERIALS: Forty rotary instruments, HCM 25/0.06 and AB 25/0.06 (n=20 each) were used. The instruments were rotated in an artificial stainless steel canal with a 60° angle and a 5-mm radius of curvature (n=10) at body temperature (35°±1°C). The torsional test evaluated the torque and angle of rotation at failure of new instruments (n=10) in the portion 3 mm from the tip according to ISO 3630-1. The fractured surface of each fragment was observed by scanning electron microscopy. The data were analyzed using unpaired student's t- test, and the level of significance was set at 5%.

RESULTS

AB 25/0.06 had significantly greater number of cycles to failure than HCM 25/0.06 (P<0.05). The torsional test showed there were no significant differences in the torsional strength and angular rotation to fracture between the groups (P>0.05).

CONCLUSION

Based on this in vitro study, AB 25/0.06 instrument was more resistant to cyclic fatigue than the HCM 25/0.06 instrument, suggested that these instruments are safer than HCM 25/0.06 for the preparation of severely curved canals. However; there was no significant difference in the torsional properties of the two instruments then appear to have similar performance during constricted canal preparation.

Для цитирования: Braga Tiago, Ricci Vivan Rodrigo, Alcalde Murilo Priori, de Camargo José MaurícioParadella, Hungaro Duarte Marco Antonio. Evaluation of the Cyclic and Torsional Fatigue Resistance of Thermally Treated Hyflex CM versus Aurum Blue Nickel-titanium Rotary Instruments.. Iranian endodontic journal. 2021. DOI: 10.22037/iej.v16i2.31006.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.