DOI 10.1038/s41372-025-02369-z
Validation of a novel Bayesian predictive algorithm for detection of carbon dioxide retention using retrospective neonatal ICU data.
Аннотация
OBJECTIVE
To validate a novel Bayesian prediction algorithm (IVCO2 index) to calculate the probability of CO2 retention in neonates using existing medical device outputs.
STUDY
DESIGN
A retrospective validation study from two level IV NICUs between September 2021 and May 2023. The algorithm calculated probabilities of PaCO2 exceeding 50 mmHg (IVCO2_50) and 60 mmHg (IVCO2_60) using multimodal physiologic data. Performance was assessed through ROC analysis, range utilization, and resolution/limitation analysis.
RESULTS
Among 180 included neonates, 1092 arterial blood gas measurements were analyzed. IVCO2_50 and IVCO2_60 demonstrated excellent discriminatory performance (AUC 0.87, 95% CI 0.85-0.89 and AUC 0.90, 95% CI 0.68-0.93, respectively). The risk of elevated PaCO2 scaled linearly with increasing index quartiles. Minimum scores (<1) showed >6-fold reduction in hypercapnia risk, while maximum scores (>99) demonstrated >3-fold reduction in normocapnia risk.
CONCLUSION
The IVCO2 index accurately predicts CO2 retention in neonates, offering potential for early detection of ventilation inadequacy without additional invasive monitoring.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.