Revision and removal of stimulating electrodes following long-term therapy with the vagus nerve stimulator.

Espinosa J, Aiello M T, Naritoku D K

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

A significant concern about vagus nerve stimulation therapy has been the disposition of the spiral stimulating electrodes once treatment is considered ineffective or is no longer desired. Because the electrodes are wrapped around the vagus nerve, there is the potential for nerve injury during their removal.

METHODS

We attempted removal of the spiral stimulating electrodes from 10 patients who received long-term vagus nerve stimulation therapy for drug-resistant epilepsy. In some patients, replacement with electrodes was also performed for poorly functioning leads.

RESULTS

The mean duration of electrode implantation was 3.7+/-2.2 years (range 1.1-7.3 years). In seven patients, the old electrodes were removed completely from the nerve. No adverse events occurred intraoperatively or postoperatively.

CONCLUSIONS

Our results indicate that the spiral electrodes may be safely removed from the vagus nerve, even after the electrodes have been implanted for several years. The reversibility of lead implantation may enhance the attractiveness of vagus nerve stimulation therapy for patients with medically-intractable epilepsy.

Для цитирования: Espinosa J, Aiello M T, Naritoku D K. Revision and removal of stimulating electrodes following long-term therapy with the vagus nerve stimulator.. Surgical neurology. 1999. DOI: 10.1016/s0090-3019(99)00046-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.