MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/rpd/ncy202

Медицинская библиотека · Radiation protection dosimetry · 2019

TYPE TESTING OF LiF:Mg,Cu,P (TLD-100H) WHOLE-BODY DOSEMETERS FOR THE ASSESSMENT OF Hp(10) AND Hp(0.07).

Pereira J S, Pereira M F, Rangel S, Caldeira M, Carvalhal G, Santos L M, Cardoso J V, Alves J G

СтатьяПеревод в очереди

Аннотация

In this work, the initial results of the type testing of the LiF:Mg,Cu,P (TLD-100H) whole-body personal dosemeters are presented. An assessment of reproducibility, linearity of the response, the residual signal as a function of the dose, energy and angular dependence of the response was performed. In general, the dosemeters show good reproducibility for different dose values and a linear behaviour for a range between 0.1 and 300 mSv. The detection limits obtained are lower than 50 μSv. The system presents a good energy and angular response for different radiation qualities.

Для цитирования: Pereira J S, Pereira M F, Rangel S, Caldeira M, Carvalhal G, Santos L M, Cardoso J V, Alves J G. TYPE TESTING OF LiF:Mg,Cu,P (TLD-100H) WHOLE-BODY DOSEMETERS FOR THE ASSESSMENT OF Hp(10) AND Hp(0.07).. Radiation protection dosimetry. 2019. DOI: 10.1093/rpd/ncy202.