DOI 10.1016/j.jacc.2009.04.023

Медицинская библиотека · Journal of the American College of Cardiology · 2009

Proof of concept: hemodynamic response to long-term partial ventricular support with the synergy pocket micro-pump.

Meyns Bart, Klotz Stefan, Simon Andre, Droogne Walter, Rega Filip, Griffith Bartley, Dowling Robert, Zucker Mark J, Burkhoff Daniel

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The purpose of this study was to test the hemodynamic effects of partial ventricular support in patients with advanced heart failure.

BACKGROUND

The use of current left ventricular assist devices (VADs) that provide full circulatory support is restricted to critically ill patients because of associated risks. Smaller, less-invasive devices could expand VAD use to a larger pool of less-sick patients but would pump less blood, providing only partial support.

METHODS

The Synergy Pocket Micro-pump device (CircuLite, Inc., Saddle Brook, New Jersey) pumps approximately 3.0 l/min, is implanted (off pump) via a mini-thoracotomy, and is positioned in a right subclavicular subcutaneous pocket (like a pacemaker). The inflow cannula inserts into the left atrium; the outflow graft connects to the right subclavian artery.

RESULTS

A total of 17 patients (14 men), age 53 +/- 9 years with ejection fraction 21 +/- 6%, mean arterial pressure 73 +/- 7 mm Hg, pulmonary capillary wedge pressure 29 +/- 6 mm Hg, and cardiac index 1.9 +/- 0.4 l/min/m(2) received an implant. Duration of support ranged from 6 to 213 (median 81) days. In addition to demonstration of significant acute hemodynamic improvements in the first day of support, 9 patients underwent follow-up right heart catheterization at 10.6 +/- 6 weeks. These patients showed significant increases in arterial pressure (67 +/- 8 mm Hg vs. 80 +/- 9 mm Hg, p = 0.01) and cardiac index (2.0 +/- 0.4 l/min/m(2) vs. 2.8 +/- 0.6 l/min/m(2), p = 0.01) with large reductions in pulmonary capillary wedge pressure (30 +/- 5 mm Hg vs. 18 +/- 5 mm Hg, p = 0.001).

CONCLUSIONS

Partial support appears to interrupt the progressive hemodynamic deterioration typical of late-stage heart failure. If proven safe and durable, this device could be used in a relatively large population of patients with severe heart failure who are not sick enough to justify use of currently available full support VADs. (Safety and Performance Evaluation of CircuLite Synergy; NCT00878527).

Для цитирования: Meyns Bart, Klotz Stefan, Simon Andre, Droogne Walter, Rega Filip, Griffith Bartley, Dowling Robert, Zucker Mark J, Burkhoff Daniel. Proof of concept: hemodynamic response to long-term partial ventricular support with the synergy pocket micro-pump.. Journal of the American College of Cardiology. 2009. DOI: 10.1016/j.jacc.2009.04.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.