DOI 10.1097/j.jcrs.0000000000001394
Sensitivity and specificity of SS-OCT for detecting macular pathologies vs SD-OCT.
Аннотация
PURPOSE: To evaluate the sensitivity and specificity of swept-source optical coherence tomography (SS-OCT) biometer compared with the gold standard spectral-domain optical coherence tomography (SD-OCT) for detecting macular pathology in patients with cataract.
SETTING
Eye Centers of Tennessee, Crossville, TN.
DESIGN
Prospective, cross-sectional, observational, examiner-masked.
METHODS
The study included 132 participants aged 50 years and older, who underwent precataract surgery work-up. All participants underwent fixation check retinal scans using SS-OCT biometer (IOLMaster 700) as well as full macular scans using Cirrus SD-OCT. 3 independent masked examiners evaluated the scans if they were normal or had a suspected pathology. Different measures of diagnostic accuracy were calculated for 3 examiners.
RESULTS
True positive rate (sensitivity) ranged from 71.1% (32/45) to 79.2% (42/53), and false negative rate was between 20.8% (11/53) and 28.9% (13/45) for the 3 examiners. True negative rate (specificity) ranged from 86.8% (59/68) to 94.1% (64/68), and false positive rate was between 5.9 (4/68) and 13.2% (9/68). The fitted receiver operating characteristic area ranged from 0.83 to 0.95.
CONCLUSIONS
Using retinal SS-OCT biometer scans as a replacement of the dedicated macular SD-OCT for screening or diagnosing macular health would not be appropriate because of its low sensitivity. SS-OCT biometer may potentially fail to identify approximately one-fourth of patients who actually have the disease. Therefore, the final decision on macular health should be based on the gold standard SD-OCT scans. When full macular SD-OCT scans are not accessible, the limited retinal scan information from SS-OCT biometer may still provide useful insights into the macular health.