Accuracy and Efficacy of Impedance Cardiography as a Non-Invasive Cardiac Function Monitor.

Kim Go Eun, Kim So Yeon, Kim Seon Ju, Yun Soon Young, Jung Hwan Ho, Kang Yhen Seung, Koo Bon Nyeo

СтатьяПеревод в очереди

Аннотация

PURPOSE: The most common method of monitoring cardiac output (CO) is thermodilution using pulmonary artery catheter (PAC), but this method is associated with complications. Impedance cardiography (ICG) is a non-invasive CO monitoring technique. This study compared the accuracy and efficacy of ICG as a non-invasive cardiac function monitoring technique to those of thermodilution and arterial pressure contour.

MATERIALS AND

METHODS

Sixteen patients undergoing liver transplantation were included. Cardiac index (CI) was measured by thermodilution using PAC, arterial waveform analysis, and ICG simultaneously in each patient. Statistical analysis was performed using intraclass correlation coefficient (ICC) and Bland-Altman analysis to assess the degree of agreement.

RESULTS

The difference by thermodilution and ICG was 1.13 L/min/m², and the limits of agreement were -0.93 and 3.20 L/min/m². The difference by thermodilution and arterial pressure contour was 0.62 L/min/m², and the limits of agreement were -1.43 and 2.67 L/min/m². The difference by arterial pressure contour and ICG was 0.50 L/min/m², and the limits of agreement were -1.32 and 2.32 L/min/m². All three percentage errors exceeded the 30% limit of acceptance. Substantial agreement was observed between CI of thermodilution with PAC and ICG at preanhepatic and anhepatic phases, as well as between CI of thermodilution and arterial waveform analysis at preanhepatic phase. Others showed moderate agreement.

CONCLUSION

Although neither method was clinically equivalent to thermodilution, ICG showed more substantial correlation with thermodilution method than with arterial waveform analysis. As a non-invasive cardiac function monitor, ICG would likely require further studies in other settings.

Для цитирования: Kim Go Eun, Kim So Yeon, Kim Seon Ju, Yun Soon Young, Jung Hwan Ho, Kang Yhen Seung, Koo Bon Nyeo. Accuracy and Efficacy of Impedance Cardiography as a Non-Invasive Cardiac Function Monitor.. Yonsei medical journal. 2019. DOI: 10.3349/ymj.2019.60.8.735.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.