[Feasibility of evaluating the left ventricular function in patients with cardiovascular diseases by fully automatic quantified three-dimensional transthoracic echocardiography].

Jing B, Tian Y, Sun Y Y, Song Y

СтатьяПеревод в очереди

Аннотация

Objective

To investigate the clinical value of left ventricular function assessment in patients with cardiovascular disease by fully automatic quantified three-dimensional transthoracic echocardiography.

Methods

One hundred and ninety-seven patients with cardiac diseases were examined by three-dimensional transthoracic echocardiography from September 2017 to May 2019. Data from 61 patients with grade 1 echocardiographic image quality were used to determine the default boundary values of endocardial end-diastolic and end-systolic phases. Clinical features were analyzed based on electronic medical records. The accuracy and repeatability of this strategy was evaluated by comparing left ventricular end-diastolic volume (LVEDV), end-systolic volume (LVESV) and left ventricular ejection fraction (LVEF) measured by automated quantitative three-dimensional echocardiography and those measured by conventional manual transthoracic echocardiography, the latter served as gold standard.

Results

The levels of LVEDV, LVESV and LVEF measured by automatic three-dimensional echocardiography were positively correlated with values obtained by manual measurement(r=0.97,0.97, 0.98, 0.97, 0.97, 0.96;P<0.05). The levels of LVEDV and LVESV measured by full-automatic three-dimensional echocardiography were significantly higher than those obtained by manual three-dimensional echocardiography(all P<0.05). The classification and correlation of systolic dysfunction in patients with abnormal ventricular wall motion by automatic three-dimensional echocardiography were significantly improved after manual calibration (κ=0.74, P=0.00) as compared to without manual calibration (κ=0.63, P=0.00). The inter-observer and intra-observer variability of fully automated three-dimensional echocardiography were significantly smaller than manual three-dimensional echocardiography(both P<0.05).

Conclusion

Fully automatic quantified three-dimensional transthoracic echocardiography possesses excellent accuracy and repeatability in measuring left ventricular volume and function, and it is feasible for clinical application.

Для цитирования: Jing B, Tian Y, Sun Y Y, Song Y. [Feasibility of evaluating the left ventricular function in patients with cardiovascular diseases by fully automatic quantified three-dimensional transthoracic echocardiography].. Zhonghua xin xue guan bing za zhi. 2020. DOI: 10.3760/cma.j.cn112148-20200721-00580.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.