Analytical performance evaluation of two automated urine chemistry analysers using two levels of commercially available quality control material.

Hokamp J A, Freeman K, Harr K E

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Evaluate two point-of-care urine chemistry analysers, VetScan SA and VetLab UA using assayed, bilevel (two concentrations) urine quality control material to determine if performance is acceptable for semiquantitative clinical urine chemistry analysis.

MATERIALS AND

METHODS

Normal and abnormal urine quality control material sent to 23 veterinary practices was evaluated three times by each clinic on in-clinic automated urinalysis instruments. Accuracy, precision and clinical utility were evaluated.

RESULTS

Normal urine quality control material: Results for blood, glucose, ketones and bilirubin were 100% accurate and precise for both analysers, and pH values were accurately acidic to neutral. However, pH from VetScan SA had clinically significant negative bias. Abnormal urine quality control material: VetScan SA: blood, microalbumin and bilirubin were 100% accurate; glucose, ketones, and protein demonstrated ≤10% inaccuracy; pH demonstrated 34% inaccuracy. VetLab UA: blood, ketones and bilirubin were 100% accurate; glucose and protein demonstrated ≤10% inaccuracy; pH was 100% accurately neutral to alkaline.

CLINICAL SIGNIFICANCE: VetScan SA had marked negative pH bias versus VetLab UA resulting in clinically significant, overly acidic results. Specific gravity, nitrite, and leukocyte test pads should not be used. Both instruments had excellent performance in normal quality control material. While blood, glucose, protein and bilirubin are correctly identified as present in abnormal quality control material, exact concentrations cannot be interpreted due to imprecision. Only semiquantitative results, not numerical values implying quantification, should be reported from urine test strips.

Для цитирования: Hokamp J A, Freeman K, Harr K E. Analytical performance evaluation of two automated urine chemistry analysers using two levels of commercially available quality control material.. The Journal of small animal practice. 2023. DOI: 10.1111/jsap.13620.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.