MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jmsacl.2021.12.008

Медицинская библиотека · Journal of mass spectrometry and advances in the clinical lab · 2022

Development of a high-throughput differential mobility separation-tandem mass spectrometry (DMS-MS/MS) method for clinical urine drug testing.

Hooshfar Shirin, Tchu Simone, Yun Cassandra, Lynch Kara L

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Differential mobility separation (DMS) is an analytical technique used for rapid separation of ions and isomers based on gas phase mobility prior to entering a mass spectrometer for analysis. The entire DMS process is accomplished in fewer than 20 ms and can be used as a rapid alternative to chromatographic separation.

OBJECTIVE

The primary objective was to evaluate the utility of DMS-tandem mass spectrometry (DMS-MS/MS) as a replacement for immunoassay-based clinical toxicology testing.

METHODS

A sensitive DMS-MS/MS method was developed and validated for simultaneous identification of 33 drugs and metabolites in human urine samples. After DMS optimization, the method was validated and used to screen 56 clinical urine samples. These results were compared to results obtained by immunoassay.

RESULTS

The DMS-MS/MS method achieved limits of detection ranging from 5 to 100 ng/mL. Moreover, the total analysis time was 2 min per sample. For the method performance evaluation, DMS-MS/MS results were compared with previously obtained urine toxicology immunoassay results. DMS-MS/MS showed higher sensitivity and identified 20% more drugs in urine, which were confirmed by LC-MS/MS.

CONCLUSION

The DMS-MS/MS as applied in our lab demonstrated the capability for rapid drug screening and provided better analytical performance than immunoassay.

Для цитирования: Hooshfar Shirin, Tchu Simone, Yun Cassandra, Lynch Kara L. Development of a high-throughput differential mobility separation-tandem mass spectrometry (DMS-MS/MS) method for clinical urine drug testing.. Journal of mass spectrometry and advances in the clinical lab. 2022. DOI: 10.1016/j.jmsacl.2021.12.008.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.