Steam-like clouding observed on anterior surface of intraocular lens developed soon after implantation.

Ota Ichiro, Miyake Goichiro, Asami Tetsu, Miyake Kensaku

СтатьяПеревод в очереди

Аннотация

PURPOSE: To report our findings in three cases of a clouding of the anterior surface of an implanted intraocular lens that developed within 14 days after implantation.

OBSERVATIONS: Three eyes were implanted with the same model IOL and a steam-like clouding developed on the anterior surface of the IOL. The clouding occurred on days 4, 7, and 14 after the implantation in an area of the IOL that was in contact with the aqueous humor. The clouding was accompanied by a reduction of vision in all cases but without any other abnormalities such as inflammation. The clouding was resolved by irrigation and aspiration of the anterior chamber in 2 cases at 2 and 34 days after the onset, and a spontaneous disappearance in 1 case at 14 days after the onset of the clouding. After the disappearance of the clouding, the visual acuity improved, and there were no recurrences. The solution used to irrigate the anterior chamber was collected and examined to confirm the absence of cellular materials. Elemental analyses confirmed that sodium and chloride were the predominant ions. The IOL implanted was the XACT lens (Advanced Vision Science, USA) which is a hydrophobic acrylic IOL and is characterized by having higher water content compared to other IOLs. In addition, it is packaged in 0.9% saline to maintain the pre-hydrated condition. These aspects may be related to the cause of the clouding. This IOL was implanted in 3271 eyes in our clinic, and 3 of them (0.09%) developed this clouding.

CONCLUSIONS AND

IMPORTANCE

We report our findings in 3 eyes that developed a clouding on the anterior surface of the lens soon after implantation. The clouding was localized to the area in contact with the aqueous. The cause of the clouding was not determined.

Для цитирования: Ota Ichiro, Miyake Goichiro, Asami Tetsu, Miyake Kensaku. Steam-like clouding observed on anterior surface of intraocular lens developed soon after implantation.. American journal of ophthalmology case reports. 2018. DOI: 10.1016/j.ajoc.2018.01.049.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.