DOI 10.1080/09593330.2014.889217

Медицинская библиотека · Environmental technology · 2014

Effects of sorbent materials on the cryofocusing analysis of gaseous reduced sulphur compounds.

Kim Ki-Hyun, Pandey Sudhir Kumar, Ahn J H, Szulejko J E, Sohn Jong Ryeul

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

The relative performance of different sorbent materials employed in the cryofocusing (e.g. in cold trap (CT) unit) stage was investigated at sub-ambient temperature by the thermal desorption (TD)-gas chromatography (GC)-pulsed flame photometric detector. To this end, the TD-based calibration of five reduced sulphur compounds (RSC: H2S, CH3SH, CS2, DMS and DMDS) and SO2 was carried out via the Peltier cooling system with five types of sorbent combinations such as two single-bed (Tenax TA and Silica gel) plus three multibed types (a combination of either two from the following three sorbents: Tenax TA, Silica gel and Carbopack B). Relative performance of each of all five CT options, if evaluated in terms of response factors for each compound, demonstrated that each CT composition acts as an important criterion to distinguish detection properties between light and heavy sulphur species. Although the relative response of H2S and CH3 SH was systematically distinguishable between the CT types, that of SO2 was the most complicated to interpret. According to this study, the two CT types consisting of Carbopack B and Silica gel (CS-0.4 and CS-0.6) were the optimum choices for sulphur gas analysis in terms of basic QA parameters (sensitivity, reproducibility and linearity).

Для цитирования: Kim Ki-Hyun, Pandey Sudhir Kumar, Ahn J H, Szulejko J E, Sohn Jong Ryeul. Effects of sorbent materials on the cryofocusing analysis of gaseous reduced sulphur compounds.. Environmental technology. 2014. DOI: 10.1080/09593330.2014.889217.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.