Electrical characteristics of pacing different portions of the His bundle in bradycardia patients.

Hu Yiran, Gu Min, Hua Wei, Niu Hongxia, Li Hui, Chen Xuhua, Zhang Nixiao, Liu Xi, Zhang Shu

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

AIMS: His-bundle pacing (HBP) can be achieved in either atrial-side HBP (aHBP) or ventricular-side HBP (vHBP). The study compared the pacing parameters and electrophysiological characteristics between aHBP and vHBP in bradycardia patients.

METHODS AND

RESULTS

Fifty patients undergoing HBP implantation assisted by visualization of the tricuspid valvular annulus (TVA) were enrolled. The HBP lead position was identified by TVA angiography. Twenty-five patients were assigned to undergo aHBP and compared with 25 patients who underwent vHBP primarily in a prospective and randomized fashion. Pacing parameters and echocardiography were routinely assessed at implant and 3-month follow-up. His-bundle pacing was successfully performed in 45 patients (90% success rate with 44.4% aHBP and 55.6% vHBP). The capture threshold was lower in vHBP than aHBP at implant (vHBP: 1.1 ± 0.5 vs. aHBP: 1.4 ± 0.4 V/1.0 ms, P = 0.014) and 3-month follow-up (vHBP: 0.8 ± 0.4 vs. aHBP: 1.7 ± 0.8 V/0.4 ms, P < 0.001). The R-wave amplitude was higher in vHBP than in aHBP at implant (vHBP: 4.5 ± 1.4 vs. aHBP: 2.0 ± 0.8 mV, P < 0.001) and at 3-month follow-up (vHBP: 4.4 ± 1.5 vs. aHBP: 1.8 ± 0.7 mV, P < 0.001). No procedure-related complications and aggravation of tricuspid valve regurgitation were observed in most patients and echocardiographic assessment of cardiac function remained in the normal range in all patients during the follow-up.

CONCLUSION

This study demonstrates that vHBP features a low and stable pacing capture threshold and high R-wave amplitude, suggesting better pacing mode management and battery longevity can be achieved by HBP in the ventricular side.

Для цитирования: Hu Yiran, Gu Min, Hua Wei, Niu Hongxia, Li Hui, Chen Xuhua, Zhang Nixiao, Liu Xi, Zhang Shu. Electrical characteristics of pacing different portions of the His bundle in bradycardia patients.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2020. DOI: 10.1093/europace/euaa309.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.