Utility of droplet digital PCR for the quantitative detection of polyomavirus JC in clinical samples.
Аннотация
BACKGROUND
Quantitative PCR (qPCR) is the standard molecular method for detection of polyomavirus JC (JCPyV) DNA reactivation in serum and cerebrospinal fluid (CSF) in patients at risk of progressive multifocal leukoencephalopathy (PML). Recently, digital PCR has shown potential benefits over qPCR in viral diagnostics.
OBJECTIVE
To evaluate the performance of droplet digital PCR (ddPCR) assay in assessing JCPyV-DNA status in clinical samples of patients at risk for PML.
STUDY
DESIGN
JCPyV specific ddPCR was developed with primers/probes targeting Large T and the noncoding control region used in qPCR. The ddPCR accuracy of JCPyV-DNA quantification was investigated using serial dilutions of genomic JCPyV-DNA. The ddPCR JCPyV-DNA quantification and qPCR confirmation were performed on 150 CSF and 100 serum clinical samples.
RESULTS
Using genomic JCPyV-DNA, ddPCR was highly sensitive, repeatable and reproducible for both molecular targets. Using clinical samples, JCPyV-DNA was detected in 13% of CSF and in 50% of serum samples with limit of detection of 30 copies/ml. Among the 19 JCPyV-DNA-positive CSF detected using the ddPCR, 15 also tested positive with the qPCR. Among the 50 JCPyV-DNA-positive serum identified with ddPCR, 41 tested positive with qPCR. All the ddPCR-negative samples were negative when assessed using qPCR. Additionally, the mean JCPyV-DNA viral load obtained with ddPCR in all samples was not significantly different from that of qPCR.
CONCLUSION
The results demonstrate that ddPCR is a highly sensitive alternative for measuring JCPyV-DNA that should be considered in clinical diagnostic testing of JCPyV-DNA in patients at risk of PML and other associated diseases.