Quantifying upper limb tremor in people with multiple sclerosis using Fast Fourier Transform based analysis of wrist accelerometer signals.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Teufl Stefan, Preston Jenny, van Wijck Frederike, Stansfield Ben
Аннотация
INTRODUCTION: Tremor is a disabling symptom of Multiple Sclerosis (MS). The development of objective methods of tremor characterisation to assess intervention efficacy and disease progression is therefore important. The possibility of using a Fast Fourier Transform (FFT) method for tremor detection was explored.
METHODS: Acceleration from a wrist-worn device was analysed using FFTs to identify and characterise tremor magnitude and frequency. Processing parameters were explored to provide insight into the optimal algorithm. Participants wore a wrist tri-axial accelerometer during 9 tasks. The FAHN clinical assessment of tremor was used as the reference standard.
RESULTS: Five people with MS and tremor (57.6 ± 15.3 years, 3 F/2M) and ten disease-free controls (42.4 ± 10.9 years, 5 M/5F) took part. Using specific algorithm settings tremor identification was possible (peak frequency 3-15Hz; magnitude greater than 0.06 g; 2 s windows with 50% overlap; using 2 of 3 axes of acceleration), giving sensitivity 0.974 and specificity 0.971 (38 tremor occurrences out of 108 tasks, 1 false positive, 2 false negatives). Tremor had frequency 3.5-13.0 Hz and amplitude 0.07-2.60g.
CONCLUSIONS: Upper limb tremor in people with MS can be detected using a FFT approach based on acceleration recorded at the wrist, demonstrating the possibility of using this minimally encumbering technique within clinical practice.
METHODS: Acceleration from a wrist-worn device was analysed using FFTs to identify and characterise tremor magnitude and frequency. Processing parameters were explored to provide insight into the optimal algorithm. Participants wore a wrist tri-axial accelerometer during 9 tasks. The FAHN clinical assessment of tremor was used as the reference standard.
RESULTS: Five people with MS and tremor (57.6 ± 15.3 years, 3 F/2M) and ten disease-free controls (42.4 ± 10.9 years, 5 M/5F) took part. Using specific algorithm settings tremor identification was possible (peak frequency 3-15Hz; magnitude greater than 0.06 g; 2 s windows with 50% overlap; using 2 of 3 axes of acceleration), giving sensitivity 0.974 and specificity 0.971 (38 tremor occurrences out of 108 tasks, 1 false positive, 2 false negatives). Tremor had frequency 3.5-13.0 Hz and amplitude 0.07-2.60g.
CONCLUSIONS: Upper limb tremor in people with MS can be detected using a FFT approach based on acceleration recorded at the wrist, demonstrating the possibility of using this minimally encumbering technique within clinical practice.