MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1515/cclm-2013-0138

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2013

Technical and clinical evaluation of the VITROS® Immunodiagnostic Products 25-OH Vitamin D Total Assay--comparison with marketed automated immunoassays and a liquid chromatography-tandem mass spectrometry method.

Cavalier Etienne, Rousselle Olivier, Ferrante Nunzio, Carlisi Agnes, Le Goff Caroline, Souberbielle Jean-Claude

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The study was conducted to evaluate the technical and clinical performance of the VITROS® Immunodiagnostic Products 25-OH Vitamin D Total Assay, and compare it with the performance of five marketed automated assays and a liquid chromatography/mass spectrometry reference method (LC-MS/MS).

METHODS

Three hundred patient serum samples were used to compare the correlation of the VITROS® 25-OH Vitamin D Total Assay with both the other immunoassays and the LC-MS/MS method, using Passing-Bablok regression and Bland-Altman analyses. Concordance of the diagnosis of vitamin D status was calculated to test the agreement between the different assays. In addition, samples containing vitamin D2 were used to test the assay's ability to detect the D2 form of the vitamin.

RESULTS AND

CONCLUSIONS

These results from the VITROS® 25-OH Vitamin D Total Assay generally correlated well with those from most of the marketed immunoassays. Cross-reactivity of the D2 form was calculated as being close to 100%. Additionally, we found substantial variability in performance amongst the various assays, which suggests the need for optimisation and recalibration of commercial methods.

Для цитирования: Cavalier Etienne, Rousselle Olivier, Ferrante Nunzio, Carlisi Agnes, Le Goff Caroline, Souberbielle Jean-Claude. Technical and clinical evaluation of the VITROS® Immunodiagnostic Products 25-OH Vitamin D Total Assay--comparison with marketed automated immunoassays and a liquid chromatography-tandem mass spectrometry method.. Clinical chemistry and laboratory medicine. 2013. DOI: 10.1515/cclm-2013-0138.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.