Duodenoscope reprocessing surveillance with adenosine triphosphate testing and terminal cultures: a clinical pilot study.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Visrodia Kavel, Hanada Yuri, Pennington Kelly M, Tosh Pritish K, Topazian Mark D, Petersen Bret T

Аннотация

BACKGROUND AND AIMS: Recent reports of infectious outbreaks linked to duodenoscopes have led to proposals for duodenoscope surveillance culturing, which has inherent limitations. We aimed to assess the feasibility of real-time adenosine triphosphate (ATP) testing after manual cleaning and its ability to predict reprocessing adequacy, as determined by terminal duodenoscope cultures.

METHODS: Clinically used duodenoscopes underwent reprocessing per current guidelines. After manual cleaning, ATP samples were obtained from the elevator, within the proximal biopsy port, and by flushing of the biopsy channel. After high-level disinfection (HLD), aerobic cultures of the elevator and biopsy channel were obtained using sterile technique. Duodenoscopes with any ATP sample ≥200 relative light units underwent repeated cycles of cleaning, ATP testing, HLD, and terminal culturing.

RESULTS: Twenty clinically used duodenoscopes were included; 18 underwent a second reprocessing cycle, and 6 underwent a third reprocessing cycle because of detection of high ATP. After the initial reprocessing cycle, 12 of 20 (60%) duodenoscopes had positive culture results, most commonly yielding gram-negative bacilli (GNB, n = 11 from 9 duodenoscopes), and catalase-positive gram-positive cocci (CP-GPC, n = 7 from 7 duodenoscopes), suggesting staphylococcal organisms. Ambient environmental controls also showed GNB and CP-GPC growth. The overall sensitivity and specificity of ATP testing compared with terminal cultures were 30% and 53%, respectively.

CONCLUSIONS: ATP sampling appears to correlate poorly with terminal culture results and cannot be recommended as a surrogate for terminal cultures. The performance and interpretation of cultures remains complicated by the potential recovery of environmental contaminants.