Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography.

Khoury Rasha, Klinger Gil, Shir Yakir, Osovsky Michael, Bromiker Ruben

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To compare the efficacy and reliability of two pulse-oximeters (POx) (Masimo Radical-7 and Nellcor™ Oxymax Bedside) and evaluate the feasibility of routine ECG monitoring during delivery room transition.

STUDY

DESIGN

Prospective observational comparative study. Sixty newborns were connected simultaneously to both POxs and ECG monitor (as a gold standard for HR). Times to achieve a stable signal were compared. Heart rates were compared to simultaneous ECG.

RESULTS

A significant difference in times to stable signal was found: Mean, Median (Interquartile range) for Nellcor and Masimo, were 15, 8.5 (6-18) and 27, 12 (9-34) seconds, respectively. Compared to ECG, false bradycardia was displayed in 18 of 55 (35%) newborns by the Masimo POx and in no newborns by the Nellcor POx. Attaching the ECG monitor was feasible but consumed additional resources.

CONCLUSIONS

The time for achievement of a stable saturation reading in an uncomplicated resuscitation setting differed significantly between POxs.

Для цитирования: Khoury Rasha, Klinger Gil, Shir Yakir, Osovsky Michael, Bromiker Ruben. Monitoring oxygen saturation and heart rate during neonatal transition. comparison between two different pulse oximeters and electrocardiography.. Journal of perinatology : official journal of the California Perinatal Association. 2020. DOI: 10.1038/s41372-020-00881-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.