His bundle pacing capture threshold stability during long-term follow-up and correlation with lead slack.

Beer Dominik, Subzposh Faiz A, Colburn Shaun, Naperkowski Angela, Vijayaraman Pugazhendhi

СтатьяПеревод в очереди

Аннотация

AIMS: His bundle pacing (HBP) is the most physiologic form of pacing. Long-term HBP capture threshold stability and its relation to lead characteristics at the time of implantation have not been adequately described. The aim of this study was to characterize HB capture threshold in follow-up and to identify potential lead characteristics predictive of lead capture instability.

METHODS AND

RESULTS

Consecutive patients with successful HBP for bradycardia indications were identified from the Geisinger HBP registry. His bundle capture thresholds, baseline comorbidities, and radiographic lead slack characteristics were analysed. An increase in HB capture threshold ≥1 V above implant values at any time during follow-up was tracked. Forty-four of the 294 studied (15%) experienced HB capture threshold increase by ≥ 1 V. Threshold increase was seen early (41% by 8 weeks, 66% by 1 year). Eighteen (6%) patients required lead revision in follow-up. Abnormal slack shape was associated with a trend toward capture threshold increase [hazard ratio (HR) 2.07; 95% confidence interval (CI) 0.9-4.6; P = 0.08]. Non-perpendicular angle of lead insertion on radiography was associated with the capture threshold increase (HR 2.81, 95% CI 1.4-5.8; P < 0.01).

CONCLUSION

His bundle capture threshold remains stable in the majority (85%) of patients. Implant characteristics may predict the threshold rise. Further evaluation of the aetiology of threshold increase and design changes in lead and delivery systems may lead to chronically stable capture thresholds.

Для цитирования: Beer Dominik, Subzposh Faiz A, Colburn Shaun, Naperkowski Angela, Vijayaraman Pugazhendhi. His bundle pacing capture threshold stability during long-term follow-up and correlation with lead slack.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2021. DOI: 10.1093/europace/euaa350.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.