Assessing the reproducibility and predictive value of objective cough measurement for successful withdrawal of invasive ventilatory support in adult patients.
СтатьяНаблюдательное исследованиеМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Varón-Vega Fabio, Rincón Adriana, Giraldo-Cadavid Luis F, Tuta-Quintero Eduardo, Palacios Jonathan, Crevoisier Stephanie, Duarte Diana C, Poveda Marcela, Cucunubo Laura, Monedero Pablo
Аннотация
BACKGROUND: Utilizing clinical tests, such as objective cough measurement, can assist in predicting the success of the weaning process in critically ill patients.
METHODS: A multicenter observational analytical study was conducted within a prospective cohort of patients recruited to participate in COBRE-US. We assessed the capability of objective cough measurement to predict the success of the spontaneous breathing trial (SBT) and extubation. Intra- and inter-observer reproducibility of the cough test and was evaluated using the intraclass correlation coefficient (ICC) and Cohen's weighted kappa. We used receiver operating characteristic curves (ROC-curve) to evaluate the predictive ability of objective cough measurement.
RESULTS: We recruited 367 subjects who were receiving invasive mechanical ventilation. A total of 451 objective cough measurements and 456 SBTs were conducted. A significant association was found between objective cough measurement and successful SBT (OR: 1.68; 95% CI 1.48-1.90; p = 0.001). The predictive capability of the objective cough test for SBT success had a ROC-curve of 0.58 (95% CI: 0.56-0.61). Objective cough measurement to predict successful extubation had a ROC-curve of 0.61 (95% CI: 0.56-0.66). The intraobserver reproducibility exhibited an ICC of 0.94 (95% CI: 0.89-0.96; p < 0.001), while the interobserver reproducibility demonstrated an ICC of 0.72 (95% CI: 0.51-0.85; p < 0.001). The intraobserver agreement, assessed using Cohen's weighted kappa was 0.94 (95% CI: 0.93-0.99; p < 0.001), whereas the interobserver agreement was 0.84 (95% CI: 0.67 - 0.10; p < 0.001).
CONCLUSIONS: The objective measurement of cough using the method employed in our study demonstrates nearly perfect intra-observer reproducibility and agreement. However, its ability to predict success or failure in the weaning process is limited.
METHODS: A multicenter observational analytical study was conducted within a prospective cohort of patients recruited to participate in COBRE-US. We assessed the capability of objective cough measurement to predict the success of the spontaneous breathing trial (SBT) and extubation. Intra- and inter-observer reproducibility of the cough test and was evaluated using the intraclass correlation coefficient (ICC) and Cohen's weighted kappa. We used receiver operating characteristic curves (ROC-curve) to evaluate the predictive ability of objective cough measurement.
RESULTS: We recruited 367 subjects who were receiving invasive mechanical ventilation. A total of 451 objective cough measurements and 456 SBTs were conducted. A significant association was found between objective cough measurement and successful SBT (OR: 1.68; 95% CI 1.48-1.90; p = 0.001). The predictive capability of the objective cough test for SBT success had a ROC-curve of 0.58 (95% CI: 0.56-0.61). Objective cough measurement to predict successful extubation had a ROC-curve of 0.61 (95% CI: 0.56-0.66). The intraobserver reproducibility exhibited an ICC of 0.94 (95% CI: 0.89-0.96; p < 0.001), while the interobserver reproducibility demonstrated an ICC of 0.72 (95% CI: 0.51-0.85; p < 0.001). The intraobserver agreement, assessed using Cohen's weighted kappa was 0.94 (95% CI: 0.93-0.99; p < 0.001), whereas the interobserver agreement was 0.84 (95% CI: 0.67 - 0.10; p < 0.001).
CONCLUSIONS: The objective measurement of cough using the method employed in our study demonstrates nearly perfect intra-observer reproducibility and agreement. However, its ability to predict success or failure in the weaning process is limited.