CVIR endovascular · 2025
FEMOSEAL CLOSE: multi-centre observational study with FemoSeal™ vascular closure device following peripheral percutaneous endovascular procedures.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Gouëffic Yann, Deloose Koen, Dubosq Maxime, Zeller Thomas
Аннотация
BACKGROUND: The purpose of the FEMOSEAL CLOSE study was to assess the safety and effectiveness of FemoSeal™ vascular closure device in achieving haemostasis following transfemoral peripheral procedures in routine clinical practice.
MATERIALS AND METHODS: This prospective, European, multi-centre, single-arm, post-market clinical follow-up study enrolled patients undergoing diagnostic or interventional peripheral endovascular procedures with FemoSeal™ vascular closure device in inpatient or outpatient settings. The primary endpoint was a composite of safety and effectiveness, with effectiveness defined as cessation of arterial bleeding without adjunctive treatment, and safety as freedom from major vascular complications in the first 6 h. Secondary endpoints included: minor access-site complications and major complications 6 h to 30 days post-procedure, and time to haemostasis, ambulation and discharge. Quality of life was assessed at baseline, discharge and 30-day follow-up. Device usability was also surveyed.
RESULTS: Between December 2021 and July 2022, 230 patients were enrolled in three European centres. The primary composite endpoint was achieved in 95.1% (215/226) [95% confidence interval: 91.46-97.55] of patients. The effectiveness component was achieved in 96.9% (219/226) [95% confidence interval: 93.70-98.70], and the safety component in 95.2% (220/230) [95% confidence interval: 92.15-97.90] of patients. FemoSeal™ vascular closure device deployment failure occurred in 1.6% (4/230) of cases, with subsequent manual compression achieving haemostasis. The median time to haemostasis was 0.42 min.
CONCLUSIONS: The study device provides effective haemostasis and low rates of access-site complications up to 30 days post-procedure for patients undergoing peripheral endovascular interventions. It demonstrates good performance with rapid haemostasis.
TRIAL REGISTRATION: The study is registered at clinicaltrials.gov. (ClinicalTrials identifier: NCT05027698).
MATERIALS AND METHODS: This prospective, European, multi-centre, single-arm, post-market clinical follow-up study enrolled patients undergoing diagnostic or interventional peripheral endovascular procedures with FemoSeal™ vascular closure device in inpatient or outpatient settings. The primary endpoint was a composite of safety and effectiveness, with effectiveness defined as cessation of arterial bleeding without adjunctive treatment, and safety as freedom from major vascular complications in the first 6 h. Secondary endpoints included: minor access-site complications and major complications 6 h to 30 days post-procedure, and time to haemostasis, ambulation and discharge. Quality of life was assessed at baseline, discharge and 30-day follow-up. Device usability was also surveyed.
RESULTS: Between December 2021 and July 2022, 230 patients were enrolled in three European centres. The primary composite endpoint was achieved in 95.1% (215/226) [95% confidence interval: 91.46-97.55] of patients. The effectiveness component was achieved in 96.9% (219/226) [95% confidence interval: 93.70-98.70], and the safety component in 95.2% (220/230) [95% confidence interval: 92.15-97.90] of patients. FemoSeal™ vascular closure device deployment failure occurred in 1.6% (4/230) of cases, with subsequent manual compression achieving haemostasis. The median time to haemostasis was 0.42 min.
CONCLUSIONS: The study device provides effective haemostasis and low rates of access-site complications up to 30 days post-procedure for patients undergoing peripheral endovascular interventions. It demonstrates good performance with rapid haemostasis.
TRIAL REGISTRATION: The study is registered at clinicaltrials.gov. (ClinicalTrials identifier: NCT05027698).