MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/00006565-199710000-00003

Медицинская библиотека · Pediatric emergency care · 1997

Comparison of self-inflating bags with anesthesia bags for bag-mask ventilation in the pediatric emergency department.

Mondolfi A A, Grenier B M, Thompson J E, Bachur R G

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare bag-mask ventilation performed by emergency department (ED) personnel using anesthesia bags (AB) and self-inflating bags (SIB).

SETTING

ED in a teaching children's hospital where the AB is the device used during resuscitations.

DESIGN

Experimental study. Bag-mask ventilation was evaluated with an infant resuscitation mannequin equipped to measure airway volumes and pressures. Pediatric residents, ED nurses, and pediatric emergency medicine fellows performed bag-mask ventilation with AB and SIB and rated their confidence using each device.

MAIN OUTCOME MEASURE: Ventilation failure rates.

RESULTS

Seventy subjects participated (17 interns, 16 junior residents, 13 senior residents, 10 fellows, and 14 nurses). There were 13 failures with the AB (18.6%) versus 1 (1.4%) with the SIB (P < 0.01) [95% confidence interval: 5-29%], with a significant difference even after excluding the least experienced subjects. There was no difference in high pressure breaths delivered (SIB 19% vs AB 15%, P = 0.4) and a higher incidence of hyperventilation with the SIB (67 vs 25%, P < 0.01). While using the SIB, 19 (27%) of the subjects did not turn on the O2 flow. There was no difference in pretest confidence rating, but the posttest confidence rating was higher for the SIB (P < 0.05).

CONCLUSIONS

Compared to SIB use for bag-mask ventilation in an ED, AB use resulted in more ventilation failures, no advantage in preventing excessive airway pressures, and less confidence among operators. The SIB should be the first choice for bag-mask ventilation in the ED, with attention to maximize oxygen delivery.

Для цитирования: Mondolfi A A, Grenier B M, Thompson J E, Bachur R G. Comparison of self-inflating bags with anesthesia bags for bag-mask ventilation in the pediatric emergency department.. Pediatric emergency care. 1997. DOI: 10.1097/00006565-199710000-00003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.