Diagnosis of patent ductus arteriosus by a neonatologist with a compact, portable ultrasound machine.
Аннотация
OBJECTIVES
To conduct a pilot study assessing a neonatologist's accuracy in diagnosing patent ductus arteriosus (PDA) using compact, portable ultrasound after limited training.
STUDY
DESIGN
Prospective study of premature infants scheduled for echocardiography for suspected PDA. A neonatologist with limited training performed study exams before scheduled exams. Sensitivity and specificity were calculated, compared to the scheduled echocardiogram interpreted by a cardiologist.
RESULTS
There were 24 exams. Compared to the scheduled exam, the neonatologist's exam had sensitivity 69% (95% confidence interval (CI), 41 to 89%) and specificity 88% (95% CI, 47 to 99%). When a cardiologist interpreted the study exams, the sensitivity was 87% (95% CI, 60 to 98%) and specificity 71% (95% CI, 29 to 96%).
CONCLUSION
A neonatologist with limited training was able to detect PDA with moderate success. A more rigorous training process or real-time transmission with cardiologist interpretation may substantially improve accuracy. Institutions with experienced technicians and on-site pediatric cardiologists may not gain from intensive training of neonatologists, but hospitals where diagnosis and treatment of PDA would be delayed may benefit from such processes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.