MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/sj.jp.7211693

Медицинская библиотека · Journal of perinatology : official journal of the California Perinatal Association · 2007

Diagnosis of patent ductus arteriosus by a neonatologist with a compact, portable ultrasound machine.

Lee H C, Silverman N, Hintz S R

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

To conduct a pilot study assessing a neonatologist's accuracy in diagnosing patent ductus arteriosus (PDA) using compact, portable ultrasound after limited training.

STUDY

DESIGN

Prospective study of premature infants scheduled for echocardiography for suspected PDA. A neonatologist with limited training performed study exams before scheduled exams. Sensitivity and specificity were calculated, compared to the scheduled echocardiogram interpreted by a cardiologist.

RESULTS

There were 24 exams. Compared to the scheduled exam, the neonatologist's exam had sensitivity 69% (95% confidence interval (CI), 41 to 89%) and specificity 88% (95% CI, 47 to 99%). When a cardiologist interpreted the study exams, the sensitivity was 87% (95% CI, 60 to 98%) and specificity 71% (95% CI, 29 to 96%).

CONCLUSION

A neonatologist with limited training was able to detect PDA with moderate success. A more rigorous training process or real-time transmission with cardiologist interpretation may substantially improve accuracy. Institutions with experienced technicians and on-site pediatric cardiologists may not gain from intensive training of neonatologists, but hospitals where diagnosis and treatment of PDA would be delayed may benefit from such processes.

Для цитирования: Lee H C, Silverman N, Hintz S R. Diagnosis of patent ductus arteriosus by a neonatologist with a compact, portable ultrasound machine.. Journal of perinatology : official journal of the California Perinatal Association. 2007. DOI: 10.1038/sj.jp.7211693.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.