DOI 10.5137/1019-5149.jtn.28409-19.2
Differences in Electrophysiologic Values between Preoperative and Intraoperative Neurophysiologic Monitoring.
Аннотация
AIM: To compare preoperative and intraoperative electrophysiologic values in patients who underwent brain and spine surgeries.
MATERIAL AND
METHODS
A retrospective medical record review was conducted on patients who underwent brain and spine surgery between January 2014 and July 2018. Patients underwent preoperative electrophysiologic monitoring within a week before the operation and intraoperative monitoring during the surgery. Monitoring parameters included the onset latency (msec) and amplitude (mV) of motor evoked potential (MEP), somatosensory evoked potential (SEP), electroneuronography, and brainstem auditory evoked potential (BAEP).
RESULTS
The latency of the MEP and SEP were significantly prolonged while the amplitude was decreased during operation. Specifically, patients with abnormal findings in their preoperative assessments showed more prominent differences compared to those with normal findings. However, there was no significant difference between preoperative and intraoperative results based on which side of the hemisphere was affected or unaffected. Unlike the MEP and SEP parameters, there were no significant changes in the facial electroneuronography and BAEP parameters.
CONCLUSION
There were differences in the preoperative and intraoperative monitoring parameters. Further studies are necessary to understand the underlying mechanisms behind these changes during surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.