DOI 10.1016/j.jcv.2024.105742

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2024

Standardization and evaluation of an in-house ELISA for the detection of rabies antibody in a tertiary care centre in South India.

Thangavelu Kalpana, Daniel Hubert D-J, Kannangai Rajesh, Abraham Asha Mary, Velladurai Selva Kumar, Mammen Dr Shoba

СтатьяОценочное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Rapid fluorescent focus inhibition test (RFFIT), a neutralization-based assay for detecting rabies antibodies, is the gold standard. The National Action Plan for Dog Mediated Rabies Elimination (NAPRE) is a national program that strategizes the establishment of enzyme-linked immunosorbent assays (ELISA) to detect rabies antibodies.

OBJECTIVE

We developed an in-house ELISA to screen for rabies antibodies using rabies vaccine antigen to study vaccine response among health care workers (HCWs) who received pre-exposure prophylaxis and a few animal bite victims who received post-exposure prophylaxis with rabies vaccine.

STUDY

DESIGN

A prospective study was carried out from April to September 2023 at the Department of Clinical Virology of a tertiary care center in South India. A total of 161 serum specimens, which included 155 serum samples from study participants and 6 samples from a reference laboratory (as controls), were obtained during the study period. Rabies antibody was determined by the in-house standardized ELISA developed using the rabies vaccine and compared with the reference assay, RFFIT. The accuracy indices of the in-house ELISA were estimated by MedCalc software (version 22.023).

RESULTS

A panel of 86 positive and 75 negative serum samples was used for evaluating the in-house standardized ELISA. The sensitivity, specificity, positive and negative predictive values of the in-house ELISA were 98.8 %, 100 %, 100 %, and 98.7 % respectively. The accuracy of the in-house ELISA is 99.4 %.

CONCLUSION

ELISA can be a practically feasible and less expensive assay compared to RFFIT which is a cumbersome procedure with a long turn-around time of 3-4 days.

Для цитирования: Thangavelu Kalpana, Daniel Hubert D-J, Kannangai Rajesh, Abraham Asha Mary, Velladurai Selva Kumar, Mammen Dr Shoba. Standardization and evaluation of an in-house ELISA for the detection of rabies antibody in a tertiary care centre in South India.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2024. DOI: 10.1016/j.jcv.2024.105742.