Archives of ophthalmology (Chicago, Ill. : 1960) · 2009
In vitro sealing of clear corneal cataract incisions with a novel biodendrimer adhesive.
СтатьяПоддержка NIH: внешняя программаНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Johnson C Starck, Wathier Michel, Grinstaff Mark, Kim Terry
Аннотация
OBJECTIVE: To determine if a biodendrimer adhesive will seal a clear corneal cataract incision.
DESIGN: An experimental study in which 2.75-mm clear corneal cataract incisions were made in 8 human donor eyes. The corneas were mounted on an artificial anterior chamber. Leaking pressure was determined in 6 corneas. These corneas were then treated with adhesive and leaking pressure was again measured in 4 of them. India ink was then applied to the 2 remaining treated and the 2 untreated corneas. Chamber pressure was cycled between 100 and 0 mm Hg. Optical coherence tomography was used to visualize the wound dynamics of a ninth cornea treated with adhesive, mounted, and pressure cycled in a similar fashion.
RESULTS: The mean (SD) leaking pressure was 77 (14) mm Hg for the nonsealed wounds and 142 (22) mm Hg for the adhesive-sealed wounds. india ink entered the nonsealed wounds and anterior chamber when the intraocular pressure was cyclically raised and lowered, whereas no india ink entered the adhesive-sealed wounds. The optical coherence tomography-visualized corneal wound did gape under pressure cycling, but the adhesive remained intact and stretched to conform to the wound.
CONCLUSIONS: Biodendrimer adhesives may be used to seal cataract wounds to prevent leakage and influx of fluid.
DESIGN: An experimental study in which 2.75-mm clear corneal cataract incisions were made in 8 human donor eyes. The corneas were mounted on an artificial anterior chamber. Leaking pressure was determined in 6 corneas. These corneas were then treated with adhesive and leaking pressure was again measured in 4 of them. India ink was then applied to the 2 remaining treated and the 2 untreated corneas. Chamber pressure was cycled between 100 and 0 mm Hg. Optical coherence tomography was used to visualize the wound dynamics of a ninth cornea treated with adhesive, mounted, and pressure cycled in a similar fashion.
RESULTS: The mean (SD) leaking pressure was 77 (14) mm Hg for the nonsealed wounds and 142 (22) mm Hg for the adhesive-sealed wounds. india ink entered the nonsealed wounds and anterior chamber when the intraocular pressure was cyclically raised and lowered, whereas no india ink entered the adhesive-sealed wounds. The optical coherence tomography-visualized corneal wound did gape under pressure cycling, but the adhesive remained intact and stretched to conform to the wound.