Clinical chemistry and laboratory medicine · 2008
Long-term stability of endogenous B-type natriuretic peptide after storage at -20 degrees C or -80 degrees C.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pereira Marta, Azevedo Ana, Severo Milton, Barros Henrique
Аннотация
BACKGROUND: B-type natriuretic peptide (BNP) is a biomarker used in the investigation of heart failure and other cardiovascular diseases. Our aim was to analyze the degradation of BNP in plasma samples kept at -20 degrees C or at -80 degrees C after 3 months or 1 year of storage.
METHODS: Using a Biosite assay-Triage BNP test, we evaluated 20 subjects with plasma BNP concentrations ranging from normal to high. Plasma aliquots were separated in five eppendorfs, one for immediate measurement after centrifugation, two for storage at -20 degrees C and two at -80 degrees C. We measured BNP after 3 months and 1 year. The significance level was set at p<0.05 (one-tailed test).
RESULTS: There was a significant decrease of 1% per month in BNP concentration. However, there was a significant interaction between time and group (high vs. low BNP). Low BNP values did not change over time, while high BNP decreased by 2% per month. The decrease over time did not depend on the storage temperature.
CONCLUSIONS: The storage of the samples at -20 degrees C or at -80 degrees C did not prevent BNP degradation, especially for high BNP values. For low BNP values, we did not observe a significant decay at either 3 months or 1 year. These results are valid for the Biosite assay-Triage BNP test.
METHODS: Using a Biosite assay-Triage BNP test, we evaluated 20 subjects with plasma BNP concentrations ranging from normal to high. Plasma aliquots were separated in five eppendorfs, one for immediate measurement after centrifugation, two for storage at -20 degrees C and two at -80 degrees C. We measured BNP after 3 months and 1 year. The significance level was set at p<0.05 (one-tailed test).
RESULTS: There was a significant decrease of 1% per month in BNP concentration. However, there was a significant interaction between time and group (high vs. low BNP). Low BNP values did not change over time, while high BNP decreased by 2% per month. The decrease over time did not depend on the storage temperature.
CONCLUSIONS: The storage of the samples at -20 degrees C or at -80 degrees C did not prevent BNP degradation, especially for high BNP values. For low BNP values, we did not observe a significant decay at either 3 months or 1 year. These results are valid for the Biosite assay-Triage BNP test.