Long-term stability of endogenous B-type natriuretic peptide after storage at -20 degrees C or -80 degrees C.

Pereira Marta, Azevedo Ana, Severo Milton, Barros Henrique

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

B-type natriuretic peptide (BNP) is a biomarker used in the investigation of heart failure and other cardiovascular diseases. Our aim was to analyze the degradation of BNP in plasma samples kept at -20 degrees C or at -80 degrees C after 3 months or 1 year of storage.

METHODS

Using a Biosite assay-Triage BNP test, we evaluated 20 subjects with plasma BNP concentrations ranging from normal to high. Plasma aliquots were separated in five eppendorfs, one for immediate measurement after centrifugation, two for storage at -20 degrees C and two at -80 degrees C. We measured BNP after 3 months and 1 year. The significance level was set at p<0.05 (one-tailed test).

RESULTS

There was a significant decrease of 1% per month in BNP concentration. However, there was a significant interaction between time and group (high vs. low BNP). Low BNP values did not change over time, while high BNP decreased by 2% per month. The decrease over time did not depend on the storage temperature.

CONCLUSIONS

The storage of the samples at -20 degrees C or at -80 degrees C did not prevent BNP degradation, especially for high BNP values. For low BNP values, we did not observe a significant decay at either 3 months or 1 year. These results are valid for the Biosite assay-Triage BNP test.

Для цитирования: Pereira Marta, Azevedo Ana, Severo Milton, Barros Henrique. Long-term stability of endogenous B-type natriuretic peptide after storage at -20 degrees C or -80 degrees C.. Clinical chemistry and laboratory medicine. 2008. DOI: 10.1515/cclm.2008.223.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.