DOI 10.1002/ccd.20795

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2006

Percutaneous closure of prosthetic paravalvular leaks: case series and review.

Pate Gordon E, Al Zubaidi Abdul, Chandavimol Mann, Thompson Christopher R, Munt Bradley I, Webb John G

СтатьяОбзорПеревод в очереди

Аннотация

BACKGROUND

Paravalvular leaks (PVLs) are a well-recognized complication of prosthetic valve replacement. Most are asymptomatic and benign, but some may cause symptoms due to a large regurgitant volume or hemolysis. Medical therapy is palliative, while reoperation carries significant morbidity and mortality. Percutaneous transcatheter closure techniques, now routinely applied in the management of pathological cardiac and vascular communications, may be adaptable to PVL closure, potentially offer symptomatic relief.

METHODS

We reviewed our experience with attempted percutaneous closure of PVLs, using data from medical and procedural records.

RESULTS

Between 2001 and 2004, 14 procedures were performed in 10 patients, all under general anesthesia, with transesophageal and radiographic guidance. Mitral (9) and aortic (1) valve replacements were involved, both mechanical and bioprosthetic. A variety of devices were used, including atrial septal occluders, patent ductus arteriosus occluders, and coils (all of label use). Six had a single procedure, which was technically successful in four: in two, the PVL could not be crossed. Four underwent a second procedure, which was technically successful in three; in one the previously deployed device was dislodged necessitating urgent, but ultimately uneventful, surgical removal and leak repair. One patient had transient severe hemolysis, which resolved after 1 week. At 1-year follow-up (9/10 pts) three had died, five had sustained symptomatic improvement while 1 patient with a residual leak still required regular blood transfusions.

CONCLUSIONS

Percutaneous closure of PVLs is time-consuming but feasible in selected patients, with a reasonable degree of technical and clinical success. A second procedure may be necessary and a variety of complications can occur.

Для цитирования: Pate Gordon E, Al Zubaidi Abdul, Chandavimol Mann, Thompson Christopher R, Munt Bradley I, Webb John G. Percutaneous closure of prosthetic paravalvular leaks: case series and review.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2006. DOI: 10.1002/ccd.20795.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.