MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/ajcp/aqab121

Медицинская библиотека · American journal of clinical pathology · 2022

Evaluation of Cytologic Sample Preparations for Compatibility With Nucleic Acid Analysis.

Sorber Laure, Claes Bart, Zwaenepoel Karen, Van Dorst Bieke, De Winne Koen, Fransen Els, Wener Reinier, Lapperre Therese, Lardon Filip, Pauwels Patrick

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

In this study, the influence of several key elements of the cytologic sample workflow on DNA and RNA content was evaluated.

METHODS

The A549 cell line, patient-derived organoids, and pleural effusions were used to investigate the effect of (1) several collection media and delayed time to processing; (2) cytology specimens; (3) cytologic staining; and (4) formalin-fixed, paraffin-embedded (FFPE) cell block processing on nucleic acid quality and quantity as determined by fragment analyzer, Qubit analysis (Thermo Fisher Scientific), and quantitative polymerase chain reaction-based analysis on the Idylla platform (Biocartis).

RESULTS

Alcohol-based collection media (CytoRich Red [Thermo Fisher Scientific] and EtOH95%) displayed high DNA and RNA preservation capacity, while phosphate-buffered saline and, to a lesser extent, formalin were associated with high RNA quality. Cytospin and smear cytology specimens were subject to DNA and RNA loss. Cytologic staining had no further impact on sample quality, hence destaining is not necessary. Both H&E-stained and unstained FFPE sections are compatible with nucleic acid analysis, despite a strong decrease in DNA and RNA quality.

CONCLUSIONS

Although several key elements of the cytologic sample workflow have an influence on DNA and RNA quality and quantity, the selection of these elements is also dependent on the downstream (ancillary) testing methods.

Для цитирования: Sorber Laure, Claes Bart, Zwaenepoel Karen, Van Dorst Bieke, De Winne Koen, Fransen Els, Wener Reinier, Lapperre Therese, Lardon Filip, Pauwels Patrick. Evaluation of Cytologic Sample Preparations for Compatibility With Nucleic Acid Analysis.. American journal of clinical pathology. 2022. DOI: 10.1093/ajcp/aqab121.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.