Accurate neonatal heart rate monitoring using a new wireless, cap mounted device.

Henry Caroline, Shipley Lara, Ward Carole, Mirahmadi Siavash, Liu Chong, Morgan Steve, Crowe John, Carpenter James, Hayes-Gill Barrie, Sharkey Don

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIM: A device for newborn heart rate (HR) monitoring at birth that is compatible with delayed cord clamping and minimises hypothermia risk could have advantages over current approaches. We evaluated a wireless, cap mounted device (fhPPG) for monitoring neonatal HR.

METHODS

A total of 52 infants on the neonatal intensive care unit (NICU) and immediately following birth by elective caesarean section (ECS) were recruited. HR was monitored by electrocardiogram (ECG), pulse oximetry (PO) and the fhPPG device. Success rate, accuracy and time to output HR were compared with ECG as the gold standard. Standardised simulated data assessed the fhPPG algorithm accuracy.

RESULTS

Compared to ECG HR, the median bias (and 95% limits of agreement) for the NICU was fhPPG -0.6 (-5.6, 4.9) vs PO -0.3 (-6.3, 6.2) bpm, and ECS phase fhPPG -0.5 (-8.7, 7.7) vs PO -0.1 (-7.6, 7.1) bpm. In both settings, fhPPG and PO correlated with paired ECG HRs (both R2  = 0.89). The fhPPG HR algorithm during simulations demonstrated a near-linear correlation (n = 1266, R2  = 0.99).

CONCLUSION

Monitoring infants in the NICU and following ECS using a wireless, cap mounted device provides accurate HR measurements. This alternative approach could confer advantages compared with current methods of HR assessment and warrants further evaluation at birth.

Для цитирования: Henry Caroline, Shipley Lara, Ward Carole, Mirahmadi Siavash, Liu Chong, Morgan Steve, Crowe John, Carpenter James, Hayes-Gill Barrie, Sharkey Don. Accurate neonatal heart rate monitoring using a new wireless, cap mounted device.. Acta paediatrica (Oslo, Norway : 1992). 2020. DOI: 10.1111/apa.15303.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.