Rapid phenotype hemoglobin screening by high-resolution mass spectrometry on intact proteins.
Аннотация
BACKGROUND
Given the excellent performance of modern mass spectrometers, their clinical application for the analysis of macromolecules is a growing field of interest. This principle is explored by hemoglobin analysis, which is a representative example by its molecular weight and clinical relevance in e.g. screening programs for thalassemia and hemoglobin variants. Considering its abundance and cellular containment, pre-analysis is significantly reduced allowing for essential rapid acquisitions.
METHODS
By parallel analysis of routine diagnostics for hemoglobin variants and thalassemia, we acquired samples of adults who were consented for hemoglobinopathy screening in our clinical laboratory. The pre-analytical process comprised of red cell lysis only; without further digestion and purification steps, the samples were directly injected in an electrospray ionization quadrupole time-of-flight setup and the intact proteins were analyzed by flow injection analysis. After optimization of process parameters, the deconvoluted mass spectra revealed the presence of α- and β-globulins. The reference ranges for the average mass of both globulins and their intensity ratio (α/β-ratio) were deduced from a disease-free subgroup and patients with a hemoglobinopathy were compared.
RESULTS
The α/β-ratio is a poor marker for thalassemia patients, yet deviant α/β-ratios are found for patients with a hemoglobin variant. Mass deviations down to 1Da can be resolved; even if the patient suffers from a heterozygotic disorder, the average mass is found outside the established reference interval.
CONCLUSIONS
Although subjects with mild thalassemia were not detected, all patients with a hemoglobin variant were resolved by top-down mass spectrometry using the average globulin mass and the α/β-ratio as screening parameters.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.