MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijcard.2020.08.098

Медицинская библиотека · International journal of cardiology · 2020

Assessment of injury current during leadless pacemaker implantation.

Breitenstein Alexander, Saguner Ardan M, Gasperetti Alessio, Hofer Daniel, Koepfli Pascal, Regoli François, Steffel Jan

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Leadless pacemakers are an established treatment option for bradyarrhythmias. Similar to conventional transvenous pacemakers, satisfying pacing values during implantation are targeted for optimal long-term device function. The objective is to investigate the role of a local injury current (IC) in leadless pacemaker implantations.

METHOD: The IC, sensing value, capture threshold and impedance were collected in 30 consecutive patients receiving a leadless pacemaker.

RESULTS

39 EGMs were recorded from 30 patients (including 9 device repositions). An IC was detected in 15 cases (38%). At implantation, the presence of an IC was associated with a significantly lower sensing (7.1 ± 3.7 mV vs 12.0 ± 4.0 mV; P = 0.004) and a higher capture threshold (median threshold 1.13 V at 0.24 ms [0.50-2.00] vs 0.50 V at 0.24 ms [0.25-0.75]; P = 0.002) and with a 26 fold higher likelihood of device repositioning compared to the absence of an IC (OR 26.3 [2.79-248], P < 0.001). Patients with an IC in their final implant position had a lower sensing (9.3 ± 4.4 mV vs 13.6 ± 4.7 mV at implantation, P = 0.04), while the initially similar capture threshold was lower after 24 h in the IC group. After 2 weeks, all parameters were similar between the two groups.

CONCLUSIONS

Our study shows that an IC can readily be observed during leadless pacemaker implantation associated with a lower sensing and a higher capture threshold at implantation but with similar to even better values during follow-up.

Для цитирования: Breitenstein Alexander, Saguner Ardan M, Gasperetti Alessio, Hofer Daniel, Koepfli Pascal, Regoli François, Steffel Jan. Assessment of injury current during leadless pacemaker implantation.. International journal of cardiology. 2020. DOI: 10.1016/j.ijcard.2020.08.098.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.