MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1600-0501.2011.02411.x

Медицинская библиотека · Clinical oral implants research · 2012

Fractured zirconia implants and related implant designs: scanning electron microscopy analysis.

Osman Reham B, Ma Sunyoung, Duncan Warwick, De Silva Rohana K, Siddiqi Allauddin, Swain Michael V

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

Two fractured one-piece experimental (commercially unavailable) zirconia implants were analyzed using scanning electron microscope (SEM) analysis to identify failure origins and aid in understanding the failure mechanisms. Modifications to the zirconia implant design are suggested to minimize such fracture incidences.

MATERIALS AND

METHODS

Two zirconia implants fractured during the final torquing in the maxillary ridge using the prescribed hand torque wrench. The implants were subsequently retrieved and prepared for optical and SEM evaluation. Critical attention was given to the fractography (crack morphology) of the fractured implants to identify the fracture origin.

RESULTS

Events related to initiation and propagation of the crack front could be detected from the morphology of the fractured surfaces. Unfavorable torque and bending forces applied on the implant during surgical placement and the inherent flaws in the material may have resulted in crack initiation and implant failure.

CONCLUSIONS

Caution must be exercised when placing zirconia implants in dense bone sites. Modification of surgical protocols for the intended implant site may be necessary. Improvement in design features specific to zirconia implants, and strict quality control during manufacture is essential to minimize the likelihood of fracture.

Для цитирования: Osman Reham B, Ma Sunyoung, Duncan Warwick, De Silva Rohana K, Siddiqi Allauddin, Swain Michael V. Fractured zirconia implants and related implant designs: scanning electron microscopy analysis.. Clinical oral implants research. 2012. DOI: 10.1111/j.1600-0501.2011.02411.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.