DOI 10.1007/bf03005381

Медицинская библиотека · Canadian journal of anaesthesia = Journal canadien d'anesthesie · 1989

Malignant hyperthermia and the clean machine.

McGraw T T, Keon T P

СтатьяПеревод в очереди

Аннотация

Following use with halothane, ten anaesthestic machines were sampled using infrared analysis for halothane contamination. Baseline measurements of halothane were made in the room and at the machine's common gas outlet. Five per cent halothane with four litres per minute oxygen flow was delivered for ten minutes into a scavenged breathing circuit. Halothane was then discontinued, an oxygen flow rate of 12 litres per minute was begun, and continuous measurements were made until the halothane concentration became undetectable. Baseline measurements of the rooms and anaesthestic machines ranged from 0 to 0.8 parts per million. Following the oxygen flow, the halothane concentration decreased to undetectable levels within six minutes in all ten machines.

Для цитирования: McGraw T T, Keon T P. Malignant hyperthermia and the clean machine.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 1989. DOI: 10.1007/bf03005381.