Interventional atrial septal defect closure using a totally bioresorbable occluder matrix: development and preclinical evaluation of the BioSTAR device.

Jux Christian, Bertram Harald, Wohlsein Peter, Bruegmann Michael, Paul Thomas

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

We sought to test the hypothesis that interventional atrial septal defect (ASD) closure can be performed safely and effectively using a bioresorbable occluder matrix.

BACKGROUND

The ideal septal occluder scaffold should promote the healthiest and most complete healing response while eventually facilitating the full resorption of the material and leaving "native" tissue behind, thus minimizing the potential for future complications from chronic foreign body and maintaining the possibility for later unobstructed transseptal access to the left atrium.

METHODS

The STARFlex occluders (NMT Medical Inc., Boston, Massachusetts) were modified by substituting the conventional polyester fabric for a bioengineered, acellular type-I collagen matrix derived from porcine submucosa with a heparin-coated surface (BioSTAR occluder, NMT Medical Inc.). Comparative transcatheter closure of ASDs was performed in young sheep (n = 36). Gross pathology and histopathology were obtained after follow-up periods ranging from 7 days to 2 years.

RESULTS

The STARFlex (control) devices were encapsulated time-dependently by ingrown fibrous tissue. Histology showed a mild but chronically persisting foreign body reaction. By contrast, BioSTAR devices exhibited a mild-to-moderate transient cellular immune response. Heparin coating of the BioSTAR surface improved the biocompatibility of the device by reducing surface thrombogencity. A remodeling process of the collagen scaffold, starting after 30 days in vivo, resulted in the full replacement of the matrix by host tissue after 2 years of follow-up.

CONCLUSIONS

The BioSTAR device is the first septal occluder with a totally bioresorbable matrix that is fully replaced by host tissue during the healing process. The promising results of this study support testing of the BioSTAR device in clinical trials.

Для цитирования: Jux Christian, Bertram Harald, Wohlsein Peter, Bruegmann Michael, Paul Thomas. Interventional atrial septal defect closure using a totally bioresorbable occluder matrix: development and preclinical evaluation of the BioSTAR device.. Journal of the American College of Cardiology. 2006. DOI: 10.1016/j.jacc.2006.02.057.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.