m+Medelevator.RU
← Поиск публикаций
Intensive care medicine · 2006

Airway humidification with a heat and moisture exchanger in mechanically ventilated neonates : a preliminary evaluation.

Оценочное исследованиеСтатьяРандомизированное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Fassassi Mikaïla, Michel Fabrice, Thomachot Laurent, Nicaise Claire, Vialet Renaud, Jammes Yves, Lagier Pierre, Martin Claude

Аннотация

OBJECTIVE: We set out to evaluate the efficacy of passive inspiratory gas conditioning in mechanically ventilated neonates and compared it with that of a heated humidifier (HH).

DESIGN: Prospective, randomized, controlled study.

SETTING: Neonatal and pediatric intensive care unit.

PATIENTS: Fourteen mechanically ventilated neonates nursed in incubators.

INTERVENTIONS: The HH was set to deliver a temperature of 37 degrees C and an absolute humidity of 40 mgH(2)O/l at the incubator entrance. Inspired temperature (T degrees ) and absolute humidity (AH) were measured by the psychometric method, transpulmonary pressure (Tpres) by means of a differential pressure transducer. Measurements were performed at 5, 10, and 15 min.

MEASUREMENTS AND RESULTS: The values of T degrees were significantly higher using the HH (34.6+/-1.6 degrees C) than the heat and moisture exchanger (HME) (33.8+/2.3, p<0.001). The values of AH were significantly higher using the HH (36.6+/-2.5 mgH(2)O/l) than the HME (32.4+/-2.8 mgH(2)O/l, p<0.001). No significant changes were observed in transpulmonary pressure. A significant positive correlation was observed between incubator temperature and the temperature delivered by the HH (R(2)=0.61, p<0.001).

CONCLUSIONS: The use of HMEs in neonates made it possible to achieve an absolute humidity of 28 mgH(2)O/l or more and a temperature of 30 degrees C or more. Higher values are obtained using a HH.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты