DOI 10.1089/dia.2007.0257

Медицинская библиотека · Diabetes technology & therapeutics · 2008

Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.

Karon Brad S, Griesmann Laurie, Scott Renee, Bryant Sandra C, Dubois Jeffrey A, Shirey Terry L, Presti Steven, Santrach Paula J

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Most glucose meter comparisons to date have focused on performance specifications likely to impact subcutaneous dosing of insulin. We evaluated four hospital-based glucose meter technologies for accuracy, precision, and analytical interferences likely to be encountered in critically ill patients, with the goal of identifying and discriminating glucose meter performance specifications likely to impact intensive intravenous insulin dosing.

METHODS

Precision, both within-run and day-to-day, was evaluated on all four glucose meters. Accuracy (bias) of the meters and analytical interference were evaluated by comparing results obtained on whole blood specimens to plasma samples obtained from these whole blood specimens run on a hexokinase reference method.

RESULTS

Precision was acceptable and differed little between meters. There were significant differences in the degree to which the meters correlated with the reference hexokinase method. Ascorbic acid showed significant interference with three of the four meters. Hematocrit also affected the correlation between whole blood and plasma hexokinase glucose on three of the four glucose meters tested, with the magnitude of this interference also varying by glucose meter technology.

CONCLUSIONS

Correlation to plasma hexokinase values and hematocrit interference are the main variables that differentiate glucose meters. Meters that correlate with plasma glucose measured by a reference method over a wide range of glucose concentrations and minimize the effects of hematocrit will allow better glycemic control for critically ill patients.

Для цитирования: Karon Brad S, Griesmann Laurie, Scott Renee, Bryant Sandra C, Dubois Jeffrey A, Shirey Terry L, Presti Steven, Santrach Paula J. Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.. Diabetes technology & therapeutics. 2008. DOI: 10.1089/dia.2007.0257.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.