Delivery sheath tear after modification for ASD closure.

Hoyer Mark H

Клинический случайСтатьяПеревод в очереди

Аннотация

During transcatheter closure of an atrial septal defect with insufficient aortic rim, a standard delivery sheath was modified by cutting a bevel at the distal tip to improve device orientation. The sheath split longitudinally when attempting to recapture the closure device. Troubleshooting allowed a device to be implanted successfully. Pitfalls regarding our sheath modification and methods to overcome prolapse of the left atrial disk into the right atrium are discussed.

Для цитирования: Hoyer Mark H. Delivery sheath tear after modification for ASD closure.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2006. DOI: 10.1002/ccd.20767.