Does helmet CPAP reduce cerebral blood flow and volume by comparison with Infant Flow driver CPAP in preterm neonates?

Zaramella Patrizia, Freato Federica, Grazzina Nicoletta, Saraceni Elisabetta, Vianello Andrea, Chiandetti Lino

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

We compared neonatal helmet continuous positive airway pressure (CPAP) and the conventional nasal Infant Flow driver (IFD) CPAP in the noninvasive assessment of absolute cerebral blood flow (CBF) and relative cerebral blood volume changes (DeltaCBV) by near-infrared spectroscopy.

DESIGN AND

SETTING

A randomized crossover study in a tertiary referral NICU.

PATIENTS AND INTERVENTIONS: Assessment of CBF and DeltaCBV in 17 very low birth weight infants with respiratory distress (median age 5 days) treated with two CPAP devices at a continuous distending pressure of 4 mbar.

MEASUREMENTS AND

RESULTS

Neonates were studied for two consecutive 60-min periods with helmet CPAP and with IFD CPAP. Basal chromophore traces enabled DeltaCBV changes to be calculated. CBF was calculated in milliliters per 100 grams per minute from the saturation rise integral and rate of rise O(2)Hb-HHb. Median (range) CBF with helmet CPAP was 27.37 (9.47-48.20) vs. IFD CBF 34.74 (13.59-60.10)(p=0.049) and DeltaCBV 0.15 (0.09-0.28) with IFD and 0.13 (0.07-0.27) with helmet CPAP (NS). Using helmet and IFD CPAP, the neonates showed no difference in mean physiological parameters (transcutaneous carbon dioxide and oxygen tension, pulse oximetry saturation, heart rate, breathing rate, mean arterial blood pressure, desaturation rate, axillary temperature).

CONCLUSION

Assessing CBF and DeltaCBV measured by near-infrared spectroscopy with two CPAP devices revealed no differences in relative blood volume, but CBF was lower with helmet CPAP. Greater active vasoconstriction and/or passive capillary and/or venous vessel compression seem the most likely reason, due to a positive pressure around the head, neck, and shoulders by comparison with the airway pressure.

Для цитирования: Zaramella Patrizia, Freato Federica, Grazzina Nicoletta, Saraceni Elisabetta, Vianello Andrea, Chiandetti Lino. Does helmet CPAP reduce cerebral blood flow and volume by comparison with Infant Flow driver CPAP in preterm neonates?. Intensive care medicine. 2006. DOI: 10.1007/s00134-006-0289-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.