The best in vitro conditions for two and three parallel stenting during endovascular aneurysm repair.

Mestres Gaspar, Yugueros Xavier, Apodaka Ana, Urrea Rodrigo, Pasquadibisceglie Savino, Alomar Xavier, Riambau Vincent

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

The aim of this study is to identify which endograft-parallel stent combinations and which degree of oversizing result in the most adequate fit in a juxtarenal abdominal aneurysmal neck, when using a double or triple parallel-stent (chimney) technique.

METHODS

In vitro silicon, juxtarenal, abdominal aortic aneurysmal neck models of different diameters, with two and three side-branches (simulating both the renal and superior mesenteric arteries), were constructed. Two different endografts of three diameters each, with two or three parallel stents (of 6 mm and 6 mm; or 6 mm, 6 mm, and 8 mm) were tested (Endurant-II endograft [Medtronic Inc, Santa Rosa, Calif] with balloon-expandable BeGraft stent [Bentley InnoMed, Hechingen, Germany] and an Excluder endograft [W. L. Gore and Associates, Flagstaff, Ariz] with self-expanding Viabahn stent [W. L. Gore and Associates]), applying three endograft-oversizing degrees: recommended (15%), excessive (30%), and over-excessive (40%). After remodeling, using the kissing-balloon technique at 37°C (98.6°F), 36 endograft-stent-oversizing models were scanned by computed tomography. The area of the gutters, parallel-stent compression, and main endograft infolding were recorded.

RESULTS

Increasing oversizing (15%, 30%, and 40%) revealed a nonsignificant propensity toward smaller gutters and similar parallel-stent compression, but it significantly augmented infolding, more in three parallel-stent models (0%, 0%, 67% and 0%, 33%, 83% of cases; P = .015 and .018, for two and three parallel-stent models; n = 36) and mainly for the Excluder-Viabahn combination. The Excluder-Viabahn showed significantly smaller gutters, but with higher stent compression, than Endurant-BeGraft combinations for both two and three parallel stents (8.2 mm2, 22.6 mm2; P = .002 and 14.4 mm2, 23.3 mm2; P = .009 gutter area; and 18%, 2%; P < .001 and 15%, 2%; P = .007 relative stent area compression, respectively).

CONCLUSIONS

Better endograft stent apposition was usually attained when using 30% oversizing during two and three parallel-stent techniques. Higher oversizing was related to nonsignificant smaller gutters but higher rates of infolding. Smaller gutters, but higher stent compression and risk of infolding, were achieved with the Excluder-Viabahn than with the Endurant-BeGraft combination.

Для цитирования: Mestres Gaspar, Yugueros Xavier, Apodaka Ana, Urrea Rodrigo, Pasquadibisceglie Savino, Alomar Xavier, Riambau Vincent. The best in vitro conditions for two and three parallel stenting during endovascular aneurysm repair.. Journal of vascular surgery. 2017. DOI: 10.1016/j.jvs.2016.09.046.