Heart rhythm · 2023
Failure to defibrillate or cardiovert due to premature truncation of biphasic shocks from implantable defibrillators.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Hauser Robert G, Kapphahn-Bergs Melanie, Casey Susan A, Witt Dawn R, Sengupta Jay D
Аннотация
BACKGROUND: In 2022 and 2023, Medtronic recalled implantable defibrillators because they may deliver less than full-energy shocks. The 2022 problem truncates the second phase of the waveform (SCP-T2), resulting in ∼32-J shocks, and is mitigated by downloadable software. The 2023 malfunction truncates the first phase of the waveform, resulting in 0- to 12-J shocks due to a glassed feedthrough problem (GFT-T1) that might be avoided by programming B>AX shock polarity.
OBJECTIVE: The purpose of this study was to assess the consequences of GFT-T1 and SCP-T2 shocks in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database and to estimate the incidences of GFT-T1 and SCP-T2.
METHODS: We analyzed MAUDE reports supplemented by Medtronic data; lead failures were excluded. The incidences of SCP-T2 and GFT-T1 were estimated using USA volumes for devices with glassed feedthroughs.
RESULTS: One hundred thirty-two devices delivered truncated shocks: 27 (20.5%) were GFT-T1; 103 (78.0%) were SCP-T2; and 2 (1.5%) truncated both phases (BOTH-T1&2). Of 54 ventricular fibrillation (VF) patients, 21 (38.9%) were not defibrillated by truncated shocks: 8 (38.1%) received GFT-T1 shocks, 12 (57.1%) received SCP-T2 shocks, and 1 received a BOTH-T1&2 shock; 2 patients suffered unrelated deaths; 1 was externally rescued; 1 outcome was unknown; the others were defibrillated by subsequent shocks or terminated spontaneously. The majority of patients (79.1%) shocked for ventricular tachycardia (VT) were converted, primarily (94.1%) by SCP-T2 shocks. Estimated incidences of GFT-T1 and SCP-T2 were 0.0078%-0.0088% and 0.1062%-0.1110%.
CONCLUSION: GFT-T1 and SCP-T2 shocks can result in failure to terminate VF/VT, but they may be preventable. Although the incidences of these truncated shocks are very low, heightened surveillance is warranted.
OBJECTIVE: The purpose of this study was to assess the consequences of GFT-T1 and SCP-T2 shocks in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database and to estimate the incidences of GFT-T1 and SCP-T2.
METHODS: We analyzed MAUDE reports supplemented by Medtronic data; lead failures were excluded. The incidences of SCP-T2 and GFT-T1 were estimated using USA volumes for devices with glassed feedthroughs.
RESULTS: One hundred thirty-two devices delivered truncated shocks: 27 (20.5%) were GFT-T1; 103 (78.0%) were SCP-T2; and 2 (1.5%) truncated both phases (BOTH-T1&2). Of 54 ventricular fibrillation (VF) patients, 21 (38.9%) were not defibrillated by truncated shocks: 8 (38.1%) received GFT-T1 shocks, 12 (57.1%) received SCP-T2 shocks, and 1 received a BOTH-T1&2 shock; 2 patients suffered unrelated deaths; 1 was externally rescued; 1 outcome was unknown; the others were defibrillated by subsequent shocks or terminated spontaneously. The majority of patients (79.1%) shocked for ventricular tachycardia (VT) were converted, primarily (94.1%) by SCP-T2 shocks. Estimated incidences of GFT-T1 and SCP-T2 were 0.0078%-0.0088% and 0.1062%-0.1110%.
CONCLUSION: GFT-T1 and SCP-T2 shocks can result in failure to terminate VF/VT, but they may be preventable. Although the incidences of these truncated shocks are very low, heightened surveillance is warranted.