DOI 10.1016/j.hrthm.2023.11.006
Failure to defibrillate or cardiovert due to premature truncation of biphasic shocks from implantable defibrillators.
Аннотация
BACKGROUND
In 2022 and 2023, Medtronic recalled implantable defibrillators because they may deliver less than full-energy shocks. The 2022 problem truncates the second phase of the waveform (SCP-T2), resulting in ∼32-J shocks, and is mitigated by downloadable software. The 2023 malfunction truncates the first phase of the waveform, resulting in 0- to 12-J shocks due to a glassed feedthrough problem (GFT-T1) that might be avoided by programming B>AX shock polarity.
OBJECTIVE
The purpose of this study was to assess the consequences of GFT-T1 and SCP-T2 shocks in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database and to estimate the incidences of GFT-T1 and SCP-T2.
METHODS
We analyzed MAUDE reports supplemented by Medtronic data; lead failures were excluded. The incidences of SCP-T2 and GFT-T1 were estimated using USA volumes for devices with glassed feedthroughs.
RESULTS
One hundred thirty-two devices delivered truncated shocks: 27 (20.5%) were GFT-T1; 103 (78.0%) were SCP-T2; and 2 (1.5%) truncated both phases (BOTH-T1&2). Of 54 ventricular fibrillation (VF) patients, 21 (38.9%) were not defibrillated by truncated shocks: 8 (38.1%) received GFT-T1 shocks, 12 (57.1%) received SCP-T2 shocks, and 1 received a BOTH-T1&2 shock; 2 patients suffered unrelated deaths; 1 was externally rescued; 1 outcome was unknown; the others were defibrillated by subsequent shocks or terminated spontaneously. The majority of patients (79.1%) shocked for ventricular tachycardia (VT) were converted, primarily (94.1%) by SCP-T2 shocks. Estimated incidences of GFT-T1 and SCP-T2 were 0.0078%-0.0088% and 0.1062%-0.1110%.
CONCLUSION
GFT-T1 and SCP-T2 shocks can result in failure to terminate VF/VT, but they may be preventable. Although the incidences of these truncated shocks are very low, heightened surveillance is warranted.