Determination of vitamin K1 in medical foods by liquid chromatography with postcolumn reduction and fluorometric detection.
Аннотация
A liquid chromatographic (LC) method is described for the determination of vitamin K1 in medical foods. The sample is enzymatically digested with lipase and alpha-amylase and extracted with 1% sodium bicarbonate solution-isopropanol (1 + 1). After C18 solid-phase extraction, vitamin K1 is separated by nonaqueous reversed-phase LC, converted to the hydroquinone by postcolumn zinc reduction, and quantitated by fluorescence detection. The limit of detection is 8 pg (3 sigma), and the limit of quantitation is 27 pg (10 sigma) on column. Linear response ranged from 0.1 to 1.0 ng vitamin K1 (r= 0.9999). The mean recovery (n = 38) for all spiking levels was 101.6 +/- 2.85%. Analysis of Standard Reference Material 1846, Infant Formula, gave a mean value of 0.95 +/- 0.088 mg vitamin K/kg (K or K1?) (n = 31) with a coefficient of variation of 9.26.